Instruction Manual DualScat Ex
26
10118E/5
5
Servicing
5.1
Service schedule
Recommended servicing work:
When
Who
What
Purpose
Once a
year or as
needed
User
Check the dry
chamber
Section 5.2
Absolutely essential for main-
taining detection accuracy and
protecting the electronics.
Once a
year or as
needed
User
Replace desiccant
in sensor head
Section 5.3
Absolutely essential for main-
taining detection accuracy.
Once a
year or as
needed
User
Clean the sensor
head
Section 5.4
Absolutely essential for main-
taining detection accuracy. In-
terval depends on medium.
Once a
year or as
needed
User
Replace O-ring
60 x 3 from sensor
head and blanking
glass or plate
Section 5.5
Absolutely essential for sus-
taining proper operation.
Once a
year or as
needed
User
Recalibrate the
photometer
Section 5.6
Advisable for maintaining de-
tection accuracy. Interval de-
pends on medium.
Every 2
years
Service or-
ganization
Replace seals
Absolutely essential for sus-
taining proper operation.
Every 5
years or as
needed
Service or-
ganization
Replace the flow
cell windows
Advisable for maintaining de-
tection accuracy. Interval de-
pends on medium and duty
conditions.
Table 4: Service schedule.
5.2
Replacement of desiccant in the Ex enclosure
Never open the photometer when cold product is flowing through the pipes.
Try to do this work during a sterilization phase or when the product is at least
at room temperature.
The dry chamber contains a desiccant to protect the optics/electronics against
moisture. The desiccant must be replaced whenever a message to this effect
appears on the display.
Section 6.2