Instruction Manual DualScat Ex
10118E/5
23
4.8
Setting the limits
Note: In order to use the limits, it is necessary to configure the relay outputs
accordingly.
Section 4.7
Up to two limits, with upper and
lower thresholds, can be pro-
grammed.
If the reading reaches the
upper
threshold
, the limit is activated; it
remains activated until the read-
ing falls back below the
lower
threshold
.
On the dual-angle instrument, it is
also possible to make a setting for
each limit indicating whether it
should react to the 90° or the
25° reading.
u p p e r th r e s h o ld
lo w e r th r e s h o ld
lim it s
lim it p a s s iv e
lim it a c tiv e
t
t
re a d in g
Figure 15: Upper and lower thresholds of a limit.
One or two limits are available only if the relays are programmed accordingly.
Section 4.7
Action
Display (Example)
Remarks
1.
Access code
> ▓00000 <
If no individual access
code was set, proceed to
step 3.
2.
Enter code:
/
change number
/
change place
Access code
> ...... <
Enter your own access
code here.
3.
* SENSOR *
* CHECK *
Instrument in service
mode.
4.
2 x
* MEAS.RANGE *
* LIMITS *
For single-angle instru-
ment, proceed to step 9.
5.
> LI1 source <
Meas.value90
LI1 = 1
st
limit
LI2 = 2
nd
limit
6.
LI1 source
> Meas.value90 <
Activate editing mode.
7.
Select source:
/
LI1 source
> ... <
Enter which reading the
limit should react to.
8.
> LI1 source <
...
Confirm selection.
9.
> Upper limit1 <
2.000 NTU
10.
Upper limit1
> 2.000 NTU <
Activate editing mode.
Limits