4. Open the cover of the buttons:
Figure 8-1
Top view
5. Set the measuring range.
Set lower range value/upper range value (without pressure available) (Page 104)
Apply lower range value/upper range value (with pressure present) (Page 119)
6. Set the pressure unit.
Pressure units [01] (Page 101)
7. Set the application of your device.
8. Set the scaling points.
Lower scaling point [18] (Page 126)
Upper scaling point [19] (Page 128)
9. Set the zero point.
Adjusting zero point (gauge pressure) (Page 116)
Adjusting the zero point (differential pressure) (Page 117)
Adjusting zero point (absolute pressure) (Page 117)
10.Lock the device.
11.Enable Functional Safety (for devices with Functional Safety).
Enabling Functional Safety over device with display (Page 160)
You can find additional functions in the section Parameter assignment (Page 97)
Commissioning
8.3 Commissioning the device with display
SITRANS P320/P420 (mA/HART)
86
Operating Instructions, 06/2018, A5E44852162-AA