4. Tighten the screw for the cover over the buttons.
5. Check the tightness of the blanking plug
⑤
and cable gland
②
in accordance with the
degree of protection.
6.3
Connect the Han cable socket to the cable
WARNING
Loss of the safety required for approval by using the Han plug
You may only use the Han plug for Ex ia devices and non-hazardous devices; otherwise, the
safety required for the approval is not guaranteed.
Note
Observe the protection class of the Han plug when defining the protection class.
The contact parts for the cable socket are supplied.
For devices with a Han plug mounted on the enclosure, make the connection via the cable
socket.
Requirement
● The terminal area of the cable socket is suitable for cables with diameters ranging from 6
to 12 mm.
● These cables use stranded wires with 1 mm
2
as single conductors ("+", "-" and ground).
● You are using a crimping tool from HARTING (article number 09 99 000 0110).
Procedure
1. Slide the sleeve and the screwed joint on the cable.
2. Strip approx. 8 mm of the cable ends.
3. Crimp the contact parts on the cable ends.
4. Assemble the cable socket.
8
+
,
$
Connector pin assignment with Han 7D or
Han 8D plug or cable socket
I
A
Output current
U
H
Auxiliary power
Connecting
6.3 Connect the Han cable socket to the cable
SITRANS P320/P420 (mA/HART)
Operating Instructions, 06/2018, A5E44852162-AA
65