82
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
1
2
0°C
(32°F)
-40°C
(-104°F)
55°C
(131°F)
80°C
(176°F)
200°C
(392°F)
150°C
(302°F)
3
4
-20°C
(-68°F)
Fig. 45: Ambient temperature - process temperature, version with temperature adapter
1 Ambient temperature
2 Process temperature (depending on the seal material)
3 Maximum permissible temperature - standard
4 Limited temperature range - plastic housing and stainless steel housing, electropolished
Vibration resistance
Ʋ
Instrument housing
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ
Rod probe
1 g with 5 … 200 Hz according EN 60068-2-6 (vibration
at resonance) with rod length 50 cm (19.69 in)
Shock resistance
Ʋ
Instrument housing
100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Ʋ
Rod probe
25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with rod length 50 cm (19.69 in)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry
Ʋ M20 x 1.5
1 x cable gland M20 x 1.5 (cable: ø 6 … 12 mm), 1 x
blind plug M20 x 1.5
Ʋ
½ NPT
1 x blind plug NPT, 1 x closing cap (red) ½ NPT
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ
Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Display and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection rating
Ʋ
unassembled
IP 20
Ʋ
mounted into the sensor without cover IP 40