64
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
Code
Text mes-
sage
Cause
Rectification
F260
Error in the
instrument
settings
– Error in the calibration car-
ried out in the factory
–
Error in the EEPROM
– Exchanging the electronics
– Send instrument for repair
F261
Error in the
instrument
settings
– Error during setup
– Error when carrying out a
reset
– False signal suppression
faulty
– Carry out a reset
– Repeat setup
F264
Installation/
Setup error
– Error during setup
– Check or correct installation
and/or parameter adjust-
ment
– Check probe length
F265
Measurement
function dis-
turbed
– Sensor no longer carries
out a measurement
– Carry out a reset
– Disconnect operating volt-
age briefly
F266
Impermissi-
ble operating
voltage
– Operating voltage below
specified range
– Check electrical connection
– If necessary, increase
operating voltage
F267
No executable
sensor soft-
ware
– Sensor cannot start
– Exchanging the electronics
– Send instrument for repair
The following table shows the error codes and text messages in the
status message "
Function check
" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
C700
Simulation ac-
tive
– A simulation is active
– Finish simulation
– Wait for the automatic end
after 60 mins.
C701
Parameter
verification
–
Parameter verification was
interrupted
–
Finish parameter verifica
-
tion
The following table shows the error codes and text messages in the
status message "
Out of specification
" and provides information on
causes as well as corrective measures.
Function check
Out of specification