Code
Text mes-
sage
Cause
Rectification
S600
Impermissi-
ble electronics
temperature
– Temperature of the pro-
cessing electronics in the
non-specified section
– Check ambient temperature
– Isolate electronics
–
Use instrument with higher
temperature range
S601
Overfilling
– Level echo in the close
range not available
– Reduce level
– 100 % adjustment: Increase
value
– Check mounting socket
– Remove possible interfering
signals in the close range
–
Use coaxial probe
S602
Level with-
in the search
range, com-
pensation
echo
– Compensation echo super-
imposed by medium
– 100 % adjustment: Increase
value
S603
Impermissi-
ble operating
voltage
– Operating voltage below
specified range
– Check electrical connection
– If necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "
Maintenance
" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
M500
Error in the
delivery sta-
tus
– With the reset to delivery
status, the data could not
be restored
– Repeat reset
–
Load XML file with sensor
data into the sensor
M501
Error in the
non-active
linearization
table
–
Hardware error EEPROM
– Exchanging the electronics
– Send instrument for repair
M502
Error in the
event memory
–
Hardware error EEPROM
– Exchanging the electronics
– Send instrument for repair
M503
Meas. reliabil
-
ity too low
–
Measurement certainty is
too low for a reliable meas-
urement
– Process component or
probe contaminated or
defective
– Check installation and
process conditions
– Clean or exchange process
component or probe
Maintenance