26
PBD-51041388
SITRANS LG250 - Operating Instructions
46201-EN-150424
3
4
1
2
Fig. 19: Top view of the plug connector
1 Pin 1
2 Pin 2
3 Pin 3
4 Pin 4
Contact pin
Colour connection ca-
ble in the sensor
Terminal, electronics
module
Pin 1
Brown
5
Pin 2
White
6
Pin 3
Blue
7
Pin 4
Black
8
5.7 Supplementary electronics
To make a second measured value available for use, you can use the
supplementary electronics - additional current output.
Both current outputs are passive and need a power supply.
The additional current output (II) cannot be used in safety-instrument-
ed systems according to SIL.
1
2
+
( )
(-)
I
4...20mA
I I
4...20mA
7
8
+
( )
(-)
3
1
2
Fig. 20: Terminal compartment, double chamber housing, supplementary elec-
tronics - additional current output
1 Current output (I) - Voltage supply of the sensor and signal output (with
HART)
2 Additional current output (II) - Voltage supply and signal output (without
HART)
3 Ground terminal for connection of the cable screen
5.8 Switch-on phase
After connecting the instrument to power supply or after a voltage
recurrence, the instrument carries out a self-check for approx. 30 s:
Plug M12 x 1 for VE-
GADIS 61
Supplementary electron-
ics - Additional current
output