Remote operation
7.1 Operating via AMS Device Manager
SITRANS LR250 with FOUNDATION FIELDBUS
76
Operating Instructions, 09/2021, A5E32221411-AH
Click on
Simulation
to open the dialog window for access to:
Simulation (Input)
Allows you to simulate the sensor value which is input to the Level Transducer Block. This
tests everything between the Level Transducer Block and Output.
Note
To activate simulation via AMS Device Manager or the 375 Field Communicator, simulation
must also be set to Enabled on the device. See
Simulate Enable (4.12.)
.
•
Simulation
•
Simulation Value
•
RAMP start
•
RAMP end
•
RAMP No of steps
•
RAMP steplength