
Parameter lists
A.4 Meter setup
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
79
Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
Pipe data
40181
300
Number of paths
Product
dependent
2
R/W
Number of paths on the sensor
40149
302
Max sample frequency 15
15
Read
The maximum frequency for flow
measurement
40147
303
Sample frequency
0.5
15
HW key The frequency at which the flow is
measured
40167
304
Rn for path 1
Product
dependent
3.4E+38
m
Read
R-factor for path 1.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).
40169
305
Rn for path 2
Product
dependent
3.4E+38
m
Read
R-factor for path 2.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).
Valid only for 2-path solutions
40171
306
Inner pipe diameter
Product
dependent
0.05 ...
1.200 m
Read
Inner pipe diameter in meters.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).