
Parameter lists
A.3 Diagnostic
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
73
Modbus
Address
Parameter
Default
value
Value range
Access
level
Description
40780
163 Previous settling date
01-01-
2000
Read Previous settling date where value
of totalizer 1 was stored
40783
164 Previous totalizer 1 value
0
dd-mm-yyyy
Read Previously stored value of totalizer 1
A.3
Diagnostic
Table A- 3
Diagnostic
Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
Diagnostic
40804
500
Latest service date
01-01-
2000
dd-mm-
yyyy
R/W*
Latest service date (can also be used
for date of installation)
40808
501
Operating hours since
power up
0
3.4E+38 h Read
Total operation hours since last power
up
40807
503
Numbers of power up
0
0 ... 65535 Read
Total number of power-ups since first
power-up.
Alarm
43005
200
Fault status
One or
more of
the
following
values: 1,
2, 3, 4, ...
up to 16
Read
Fault status
1: Path 1 error
2: Path 2 error
3: Dual slope calibration error
4: Database checksum error
5: Low power warning
6: Flow overload warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval error
10: Not used
11: Not used
12: Not used
13: Not used
14: Not used
15: Not used
16: Not used
40346
202
Date of fault log reset
Last reset
date
dd-mm-
yyyy,
hh:mm:ss
Read
Date of last fault log reset
14
204
Reset the fault log and
faults
No
No, Yes
R/W
Reset the fault log and faults
15
205
Call up acknowledge
No
No, Yes
R/W
Select 'Yes' to reset active call-up