Parameter lists
A.3 Diagnostic
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
77
Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
40803
515
Controlling output A and
B
Auto
Auto,
Forced
HW key Control of output A and B is used to
stop or force the output A and/or B.
For editing, use device menu 'Service'.
Valid if pulse output is enabled
Power
43120
130
Consumed battery
capacity
e.g.
2.054239
Ah
Read
The amount of energy consumed since
last time the battery was replaced.
Valid only for battery-powered versions
40808
501
Operating hours since
power up
1105 h
3.4 E+38 Read
Total operation hours since first power
up
40807
503
Numbers of power up
e.g. 4
0 ... 65535 Read
Total number of power-ups since first
power-up.
18
540
Battery change enable
No
No, Yes
R/W
Select 'Yes' to set battery installation
date to current date and reset
remaining battery operation capacity to
maximum.
Valid only for battery-powered versions
40881
541
Battery installation date last
battery
installation
date
dd-mm-
yyyy;
hh:mm:ss
Read
Latest installation date of batteries.
Valid only for battery-powered versions
40880
542
Power supply mode
Battery
only
Battery
only,
Mains
only,
Mains with
backup
battery.
Read
The flowmeter is either powered by
battery only, mains only or by mains
with battery backup.
The type is defined by the order code
and cannot be changed.
40849
543
Actual battery capacity
25.000000 0 ... 50
R/W**
The capacity of the battery in Ah
- Single 12.5 Ah
- Dual battery pack 25.0 Ah
Valid only for battery-powered versions
40851
544
Battery alarm limit
80
0 ... 90
R/W
Present an alarm when the consumed
energy superceeds this percentage of
the battery capacity.
Valid only for battery-powered versions
* Not supported for FUS080
** Maintenance = Read only