Parameter lists
A.3 Diagnostic
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
75
Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
40283
228
Parameter checksum
fault appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
40286
229
Parameter checksum
fault disappears
e.g. 0
0 ... 65535 Read
Password to change parameters in the
flowmeter.
Valid only if alarm is enabled
7
230
Low power alarm output
enable
Yes
No, Yes
R/W
Select 'Yes' to enable active alarm on
alarm output / call-up
40289
231
Low power fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40290
232
Low power fault counter e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
40291
233
Low power fault appears e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
40294
234
Low power fault
disappears
e.g. 0
0 ... 65535 Read
Last time the fault disappeared.
Valid only if alarm is enabled
8
235
Flow overflow alarm
output enable
Yes
No, Yes
R/W
Select 'Yes' to enable active alarm on
alarm output / call-up
40297
236
Overflow fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40298
237
Overflow fault counter
e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
40299
238
Overflow fault appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
40302
239
Overflow fault
disappears
e.g. 0
0 ... 65535 Read
Last time the fault disappeared.
Valid only if alarm is enabled
9
240
Pulse A overload alarm
output enable
Yes
No, Yes
R/W
Select 'Yes' to enable active alarm on
alarm output / call-up
40305
241
Pulse A overload fault
hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40306
242
Pulse A overload fault
counter
e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
40307
243
Pulse A overload fault
appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled
40310
244
Pulse A overload fault
disappears
e.g. 0
0 ... 65535 Read
Last time the fault disappeared.
Valid only if alarm is enabled
10
245
Pulse B overload alarm
output enable
Yes
No, Yes
R/W
Select 'Yes' to enable active alarm on
alarm output / call-up
40313
246
Pulse B overload fault
hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40314
247
Pulse B overload fault
counter
e.g. 0
0 ... 65535 Read
Total number of faults.
Valid only if alarm is enabled
40315
248
Pulse B overload fault
appears
e.g. 0 h
0 ... 65535 Read
First time the fault appeared.
Valid only if alarm is enabled