LITHOSTAR MODULARIS
4 - 17
Siemens AG
SPL1-130.840.02
Page 17 of 34
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
•
To support arm serial number 0050
- Place the water supply (W/Fig. 20) in the shock wave head in the same manner it was
done during deinstallation.
-
Reattach the white teflon plate
(Tp/Fig. 18).
•
From support arm serial number 0051
- Reattach the lower part (2/Fig. 21).
- Tighten the two screws (1/Fig. 21).
•
Place the temperature sensor (T/Fig. 18) with the spiral spring from the back in the shock
wave head in the same manner it was done during deinstallation; ensure that the spiral
force of the spring pushes the temperature sensor to the front (Fig. 18).
•
Reattach the disk (3/Fig. 24) to the corrugated tubing, make sure the leads are correctly
attached.
- High voltage cable (4/Fig. 24)
- Tubing sections (5/Fig. 24)
- Other lead (6/Fig. 24).
•
Reattach the corrugated tubing with the screws.
•
Route the cables and the hoses according to Fig. 24.
•
Reattach the column cover.
•
Mount the shock wave head assembly (refer to the corresponding section).
•
Reattach all covers (refer to the corresponding section).
Fig. 21
Fig. 22
1
1
1
2