LITHOSTAR MODULARIS
4 - 19
Siemens AG
SPL1-130.840.02
Page 19 of 34
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
•
Install the water supply (W/Fig. 20) in the shock wave head support, following the same
procedure as for removal.
•
Reinstall the white teflon plate (Tp/Fig. 18).
•
Install the temperature sensor (T/Fig.20) with the spiral spring in the shock wave head
support following the same procedure as for removal in the reverse order; when doing
this, make sure that the spring force of the spiral spring is pushing forward against the
temperature sensor (Fig. 18).
•
Install the shock wave head assembly (refer to the corresponding section).
•
Reattach all covers (refer to the corresponding section).
From support arm serial number 0051
4
•
Remove the covers of the basic system unit and the shock wave head.
•
Empty the coupling circuit (refer to specific paragraph).
•
Remove the coupling bellows.
•
Remove both screws (1/Fig. 21).
•
Pull out the lower part (2/Fig. 21).
•
Remove the part with the air suction hose (1/Fig. 22).
•
Replace the air suction hose.
•
Pierce the hose two times with a needle (approx. 1mm) in the same manner the old hose
was pierced.
•
Reassemble in the opposite order.Ensure proper position of the Allen screws during
reassembly.
Fig. 25
F
E