Iso-center with ultrasound
•
If the center point of the white area is not in the isocenter (allowed deviation ± 0,5 mm
from the center):
- Note the value of D2 (with sign):
If the ball image is displayed shifted to the left of the isocenter, then the correction
value must receive a negative sign (to the right is positive sign).
- Press the
F6
key.
- Select "MUP" with the trackball.
- Press the
"Set"
key
- Press the "5" key.
- Enter the password.
- Select ok with the trackball.
- Press the "Set" key.
- With the trackball select the value for image tilt and increase it with the "Set" key.
- Set half value of D2.
Example:
D2 = 1.2 mm
⇒
enter -06 .
- Press the
key.
- Check the deviation, if the white area is not in the center of the isocenter, repeat the
process.
•
After the setting fasten the holder for the ultrasound probe to the holding point (7/Fig. 1).
With every probe check that the setting in this position is also in order. Make no changes
to the image tilt setting.
•
After the check stow the parts in the corresponding transport cases.