Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Final procedures
19-3
Mounting the stand covers
The Mammomat stand covers exist in two versions, one with separate rear side covers
and one with a single rear cover that also covers the rear sides.
Mammomat stand with separate rear side covers.
Figure 1 Mounting the stand covers on a stand with separate rear side covers.
1.
Fit the rear cover (3) to the stand and fasten it with screws (4) and contact
washers (1).
Be careful not to damage the cables at the cable outlet!
2.
Fit left and right front covers (8) to the stand and fasten with screws and contact
washers (9).
3.
Fit left and right rear side covers (5) (6) to the stand and fasten with screws and
contact washers (7).
4.
Fit inner plastic strip (10) to the front and side covers and fasten with screws (11),
eight on either side.
5.
Press on the outer plastic strip (12).
When mounting the stand covers, make sure that the screws at
the top of the stand are securely tightened and have functioning
contact washers (1). These screws are used to establish protec-
tive ground connections. Before mounting the front covers, be
sure to remove the protective strips from the metal curtain (2).
Store the strips in the holders provided on both sides of the cur-
tain.
WARNING
1
4
5
9
3
6
11
12
10
2
8
ST_COV1
7
MAM00146