Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Start-up and functional test of the IONTOMAT
12-11
Output values of constancy testing (§16 Germany)
Determine the output values of the constancy testing at approx. 30 kV with the customer´s
own constancy test body (NORMI 7 test body etc.) and enter them in the form stapled to
the acceptance test certificate.
Procedure:
1
On the control panel, select a value of approx. 30 kV and AEC.
2
Place the test body on the object table flush with the chest wall side.
3
Slide film cassette into the object table.
4
Trigger the exposure and develop the film.
5
Measure the resolution (A-D), density (1-4) and the over-radiation on the chest wall
side and enter in the form (steel balls).
6
Read off the pen dosimeter and record the dose value.
Plexiglass steps:
Woven steel wire:
1 = 40 mm
A = 100
µ
m
2 = 43 mm
B = 80
µ
m
3 = 49 mm
C = 63
µ
m
4 = 46 mm
D = 40
µ
m
NORMI 7
test body
Double detector
Radiographic system
Film cassette
Pen dosimeter
(Front view)
Pen dosimeter
A
B
C
D
1 2 3 4
(Top view)
Radiographic
system
NORMI 7
test body
Test arrangement
Steel balls