65 02 186
Mains connection and power supply
7-1
Checks before powering up the generator
On p.c. board D702:
1
Set switch S1(UZW) to "OFF" (lower position).
2
Set switch S2 (SS) to "OFF" (lower position).
3
Set switch S3 (TEST) to "ON" (upper position- output service PC).
Figure 1 P.C. board D702
Measuring the line resistance
See Figure 2
1
Remove fuses F10 (1) and F20 (2) or, when single-phase supply, fuse F10 (1).
2
Connect line resistance meter to fuse holders F10 and F20.
3
Mains supply "ON".
4
Carry out measurement.
5
Mains supply "OFF".