Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Final procedures
19-1
Service PC and measuring instruments
Recording the programmed values
Save the programmed values at menu command "SAve config.file".
Main menu:
1
Select Configuration
<↑> <↓>
< ENTER>.
2
Select SAve config.file
<↑> <↓>
< ENTER>.
3
Press <F2>.
Save data with <F2>.
NOTE!
No write protection on the diskette !
Reading the exposure counter and deleting the error memory
Read the exposure counter and delete the error memory after putting the stand into ser-
vice.
A/ Reading the exposure counter
Main menu:
1
Select Service
<↑> <↓>
< ENTER>.
2
Select Show exposure counter
<↑> <↓>
< ENTER>.
3
Enter the current counter reading in the certificate.
B/ Deleting the error memory
Main menu:
1
Select Service
<↑> <↓>
< ENTER>.
2
Select Delete error buffer
<↑> <↓>
< ENTER>.
3
Delete the error memory with <Y>.
4
Leave the program with <F10>.