
Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Checks without high voltage
10-1
Checks without high voltage
1
System "OFF".
2
On p.c. board D702 set switch S1 (UZW) to "ON" (upper position) and switch S2
(SS) to "OFF" (lower position).
3
System "ON".
4
Press the exposure buttons on the control panel. The rotating anode shall start up. At
the end of the limit time, the rotating anode is braked.
5
Release the exposure buttons.
NOTE!
If the exposure buttons are prematurely released, "Limit" will be indicated.
6
System "OFF".