Cable connections
Connecting the stand cable-harness
The high-voltage connector, alternative types
Figure 4 The high-voltage connectors, alternative types
NOTE!
The high-voltage connector (1) is factory-adjusted for correct fit. Mount the angular ca-
ble-sleeve halves and sleeve nut (Type B only). The high-voltage connector shall be in-
serted with silicone oil, silicone-rubber ring (2) and silicone-rubber disc (3). Lubricate
the silicone-rubber disc and the pins with silicon oil before mounting the disc. Make sure
that there aren’t any air bubbles between the disc and the connector.
3
1
2
MA
M001
61
Type A
3
1
2
MA
M000
93
Type B