Commissioning
3.4 Setting the Parameters
SITRANS FUH1010 Quick Start
26
Operating Instructions (Compact), 12/2008, CQO:QSG010 Revision A
4.
Trigger <ENTER> to save selection. The [Anomaly Diams] menu will appear.
$QRPDO\'LDPV
!
3LSH&RQILJ
)XOO\'HYHORSHG
①
Use this menu cell to select the pipe configuration that most accurately represents the
upstream pipe condition.
②
Use this menu cell to enter the number of pipe diameters between the upstream
configuration and the transducer installation.
5.
Trigger <Right Arrow> to select "=" sign.
6.
Trigger <Up/Down Arrows> to select number of pipe diameters.
7.
Trigger <ENTER> to save selection.
8.
To return to the top menu level, continue to trigger the <Left Arrow>.
Table 3- 3
Pipe Configuration Option List Definitions
Options
Definitions
Fully Developed
Fully developed flow, as would be expected for very long
straight pipe runs or installation downstream of a flow
condition.
1 Elbow
Single 90 degree Elbow upstream of transducer installation.
Dble Elbow+
Double out-of-plane Elbows upstream of transducer
installation.
Dble Elbow-
Double in-plane Elbows upstream of transducer installation.
Valve
Not available at this time.
Expander
Pipe expansion upstream of transducer installation.
Reducer
Pipe reduction upstream of transducer installation.
Norm Entry
Not available at this time.
Header Inlet
Header or pipe manifold upstream of transducer installation.
Intrusions
Not available at this time.
Transducer Selection Example for D1H High Precision Transducers
1.
At the [Chan/Path Setup] menu, trigger the<Down Arrow> to select [Pick/install Xdcr].
2.
Trigger <Right Arrow> then <Down Arrow> to [Transducer Model].
3.
Trigger <Right Arrow> sensor.
4.
Trigger <Down Arrow>and scroll to [1011HP-T1.]. Trigger <ENTER> to select.
5.
Trigger <Down Arrow> to select [Transducer Size].