Commissioning
3.4 Setting the Parameters
SITRANS FUH1010 Quick Start
Operating Instructions (Compact), 12/2008, CQO:QSG010 Revision A
23
3.4
Setting the Parameters
Flowmeter Programming
Note
Before creating a site select a language and then English or metric units from the Meter
Facilities menu.
Select a Meter Type
1.
After power-up, use the Magnetic Wand to trigger the <ENTER> sensor and access the
top level of the Installation Menu.
2.
Trigger the Right Arrow> sensor.
3.
Trigger <Up/Down Arrows> to select [Dual Path Flow]
4.
Trigger <ENTER> sensor to select meter type: [Channel 1 Clamp-on] appears.
5.
To select [Channel 1 Clamp-on] mode, trigger <Right Arrow> sensor.
6.
Trigger <ENTER> sensor to select mode. [Chan/Path Setup] menu will appear.
0HWHU7\SH
'XDO3DWK)ORZ
①
ENTER Sensor
③
Current Selection
②
ENTER Target Symbol Sensor
④
Top level of Installation Menu
Note
IMPORTANT: Colon appears when menu item is selected for entry.
Create a Site
1.
Trigger <Right Arrow> and then <Up/Down Arrows> to select [Create/Name Site].
2.
Trigger <Right Arrow> again to select the "?" symbol (see figure below right).