SITRANS FUH1010 Quick Start
Operating Instructions (Compact), 12/2008, CQO:QSG010 Revision A
17
Commissioning
3
3.1
Connecting Power
DANGER
Turn off main power before installing AC connections to the flowmeter. Contact with
exposed wiring may lead to fire, electric shock, or serious personal injury.
1.
Using a 1/16" Hex key, loosen the flowmeter Rear Housing Cover locking setscrew.
2.
Unscrew the Rear Housing Cover and remove.
3.
Locate power supply connector J8. Use the tables below and the part number of your unit
to determine the correct input power source. (See ID label on right side of housing.)
;
=3
/
7
1
6
.
①
Flowmeter Number
⑤
Power Supply Option
②
Transducer Type
⑥
Language Option
③
Relay Type
⑦
Safety Rating Option
④
Numeric Display
Power Supply Option Code
Power Supply P/N
User Supplied Power
S
1010X-6SS2
90-250 VAC Single Phase
ZN
1010X-6ZNS2
9-36 VDC Negative Ground
ZP
1010X-6ZPS2
9-36 VDC Positive Ground
Note
Power Supply Module P/N label is located inside the front housing cover.
4.
Pull the desired length of input power wires through flowmeter case conduit cable hole
and into flow meter case before wiring connector.