Operating manual
37
------------------------------------------------------------------
------------------------------------------------------------------
Submenu CALIBRATION
------------------------------------------------------------------
Name .... "Pyrometer PT Series"
Menu mode ............. default
1: [LAMBDA 1 CALIBRATION]
A: Reset settings to factory default
E: Set menu mode
S: Set pyrometer name
Z: End Calibration-Mode
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>CALIBRATION >
You can reset all configurations you have made to your pyrometer and
restore the factory default settings using command "A". This also applies
to data acquisition parameters, LED and buzzer.
Press command "E" if you would like to view all parameters which are
otherwise hidden. Because this list of menu items is quite long (Chapter
16), we only recommend this for advanced users. If you intend to carry
out measurements of molten metal only, it makes sense to maintain the
factory configuration. This will keep those parameters, which will not re-
quire adjustment, hidden
Command "S" enables the user to enter a short word to designate the
measurement location/task. This text can be accessed from the Main
Menu by pressing "Q".
------------------------------------------------------------------
Submenu LAMBDA 1
------------------------------------------------------------------
L1 range .... 0.0 - 1000.0 C
L1 User calibration ....... off
L1 User def. offset +0.00000
L1 User def. factor +1.00000
A: Set L1 - extended-range
B: Set L1 User-Cal. On/Off
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>CALIBRATION >LAMBDA 1 >
Use command "A" to reconfigure
the pyrometer‘s measuring range. The
range can be adjusted to be either larger or smaller than the factory-
configured range. When customizing the temperature range, make sure
that upper and lower temperatures you enter are compatible with the py-
rometer
’s actual working range.
Use key "B" for direct access to enable adjustments to the Ardoptix PT.
Summary of Contents for Ardoptix PT 140 AF1
Page 1: ...Two colour Pyrometer PT 14x 7MC3040 Operating Manual 03 2017 ...
Page 2: ......
Page 3: ...Two colour Pyrometer PT Ident Nr 106 7393 PT14x_si_en doc Stand 21 03 2017 830 hm 810 het ...
Page 8: ...s Operating manual ...
Page 56: ...s Operating manual 48 29 Dimensions ...
Page 63: ...s Operating manual 55 ...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......