s
Operating manual
31
Direct commands have an assigned key. Example: E for epsilon (ratio
correction). Submenu settings are shown in brackets. Exam-
ple
[Quotient]
19.1 Main Menu
After starting the terminal program
or after entering „H“ the main menu
will appear
------------------------------------------------------------------
>H
------------------------------------------------------------------
Mainmenu
------------------------------------------------------------------
0: [QUOTIENT] E: Quick access EPSILON
1: [LAMBDA 1] A: Quick access FILTER
2: [LAMBDA 2] T: Quick access TYPE of measure
C: [I/O]
K: [CALIBRATION]
H: Show this help-site J: Show diagnosis
W: Show ambient temperature Q: Show calibration data
X: Show measure temperatures P: Show channel parameters
------------------------------------------------------------------
19.2 View Current Configuration
Comman
d „P“ shows you how your Ardoptix PT is currently configured.
------------------------------------------------------------------
- PT 143 AF1 650-1700C - 00/00112 - Job - 21.10.13 -
- PT40SW101/0 QP 0,95/1,05um Version 01.02 10.07.13 -
------------------------------------------------------------------
Qu range .... 650.0 - 1700.0 C DISPLAY source ....... quotient
Qu epsilon ratio ...... 100.0 %
Qu check L2 rel.limit 10.00 %
Qu abs.limit 650 C @ 50.00 %
Qu linearization .......... off
Qu filter .......... 0.10 s
Qu memory type ............ off GRN.LED source ... ready-signal
GRN.LED function level/signal
Unit .................. Celsius GRN.LED delay time ... 0.00 s
Terminal assigned to ...... USB GRN.LED hold time .... 0.00 s
Autoprint ......... on (cyclic) BUZZER source ............ off
Print cycle time ..... 0.1 s
Protocol address .......... 001
Display ........... temperature
Key lock .................. off
------------------------------------------------------------------
>
At the top left, there is a list of data acquisition parameters for the two
colour reading (Quotient). The right-side column shows LED/Buzzer-
Configuration. At the bottom left you will find general settings.
Summary of Contents for Ardoptix PT 140 AF1
Page 1: ...Two colour Pyrometer PT 14x 7MC3040 Operating Manual 03 2017 ...
Page 2: ......
Page 3: ...Two colour Pyrometer PT Ident Nr 106 7393 PT14x_si_en doc Stand 21 03 2017 830 hm 810 het ...
Page 8: ...s Operating manual ...
Page 56: ...s Operating manual 48 29 Dimensions ...
Page 63: ...s Operating manual 55 ...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......