35
Operating instructions
14.5 Reset all parameters to factory settings
►
[
rES.
] select in menu advanced functions
►
Press [ENTER]
> The display shows [STOP]
►
Hold [ENTER] for 2 seconds
> Display flashes for 2 seconds
► Press [▲]
> The display shows [
EHEc.
]
►
Press [ENTER]
> The display shows the current temperature
After a reset to factory settings configure the emissivity coefficient [
EPS
]
again (→ 10 Setup) to obtain exact measuring results.
15 Operation
After connecting the supply voltage the pyrometer will be automatically initialized
and will perform a self-diagnosis. After approx. 0.5 sec the sensor is ready to
operate and the instrument runs the signal processing.
The parameters are described in chapter 10 / page 25.
15.1 Ambient temperature
The maximum permissible ambient operating temperature for the pyrometer is
65 °C. If the instrument is used in ambient temperatures above 65 °C, it must be
either cooled or shielded from excess radiant heat by means of a deflector plate.