26
Operating instructions
12 Setup
12.1 Two-colour mode
When measuring with a two-colour pyrometer, the field of view does not have to be
filled entirely by the target object. There might even be particles like dust, steam
and smoke in the viewing path that weaken the infrared radiation. LED 3 signals
when a measurement is critical.
The signal strength is calculated for this signal.
LED 3 lights up Measurement OK
LED 3 flashes
Signal strength below the set threshold - measurement is critical.
LED 3 off
The signal strength is too low, a measurement is not possible.
►
Set the pyrometer to two-colour mode (Default):
Two-colour [
Ao
] → [
Ao.S
] =
Quot.
►
Align the pyrometer to the object to be measured
Sighting hole smaller
than th epyromter field of view
Emitted
radiation
Dust, steam and
smoke in atmosphere
Target smaller than
the filed of view
and/or moves
Emitted
radiation
Emitted
radiation
Dirty lens or
sighting window
Emitted
radiation