Operating instructions
12.2 One-colour mode
►
Set the pyrometer to one-colour mode:
One-colour [
Ao
] → [
Ao.S
] =
SpEc.
►
Align the pyrometer to the object to be measured
The pyrometer uses the intensity of infrared radiation for non-contact temperature
measurements. It is necessary to configure the pyrometer to the respective
emissivity coefficient of the measuring object to obtain exact measuring results (→
15 Emissivity coefficient tables). An incorrectly set emissivity coefficient leads to
wrong temperature readings.
Set the emissivity coefficient after connecting the supply voltage or resetting the
parameters to factory settings.
►
Press [
▲or ▼]
> the value of the selected emissivity is displayed, for example [
100.0
]
►
Press [
▲or ▼] until the desired emissivity is shown
►
Press [Enter] or wait for 3 seconds
The current temperature value is displayed. The pyrometer now works with this
configured emissivity until it is changed again.
Target larger
than spot size
Best
Good
Incorrect
Target equal
to spot size
Target smaller
than spot size