
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
96
3.1.4
Discrete Output Block with ULTRAMAT, OXYMAT and CALOMAT
The input value from the discrete output function block in the AUTO mode is the setpoint which is transferred
as a cyclical use date for triggering AUTOCAL using the PROFIBUS of the optional module.
Rel.
Index
Parameter
Description
9
SP_D
Transfer value/digital value in the AUTO mode
10
OUT_D
Calculated output value of the discrete output function block.
19
FSAFE_TIME
Monitoring time in seconds. The failsafe control monitors the bus
communication, in particular the setpoints SP_D and RCAS_IN.
12
READBACK_D
The readback value contains the current state in the device (actual
value).
20
FSAFE_TYPE
Defines the reaction of the device if the used setpoint is still bad after
the monitoring time FSAVE_TIME has expired. One of three reactions
can be set:
0:
The substitute value FSAFE_VAL_D becomes effective.
1:
The last valid set value is still issued. The value has the
“Uncertain, last valid value“ status.
21
FSAFE_VAL_D
Substitute value in failure behavior and when reaction type
FSAFE_TYPE = 0 is set.
24
SIMULATE
Simulation value and status for the actual value, enabling of
simulation.
33
CHECK_BACK
not used
34
CHECK_BACK_MASK not used