ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
137
7.5.11 Function Block DO
Function Block DO (Digital Output)
Class
Sl
ot
Index
abs
Ind
ex
rel
Elem
ent
Default
Value
Object
Parameter description
Data Type
Num
ber
Store
(acc. to
profile)
cy
cli
c
re
ad
wri
te
BLOCK
5
16
0
BLOCK_OBJECT
Block characteristic
ds32
1
C
x
1
250
Reserved
250 = not used
2
2
Block Object
Block-Typ (Function)
3
2
Parent Class
Parent class: 01 = Input
4
2
Class
01 = Analog Input
5
0
DD-Reference
reserved
6
0
DD-Revision
reserved
7
0x4002
Profile
PA-profil-number assigned by PNO for
PROFIBUS - PA = 64
8
0x0300
Profile Revision
Profile 3.0
9
0
Execution Time
f0r future use
10 45
NumberOfParameters
number of parameters
11 0x054b
Address of VIEW 1
Index of view 1
12 1
Number of View Lists
Number of views
STD
17
1
0
ST_REV
Identification / device / statistic rev.-no. /
Function Block: counter which is incremented
at any change of configuration parameters
unsigned16
1
N
x
STD
18
2
TAG_DESC
Unique tag in the system which can be used
by the user
visible_string
32
S
x
x
STD
19
3
0
STRATEGY
Strategy
unsigned16
1
S
x
x
STD
20
4
0
ALERT_KEY
User can write value for alert handling
unsigned8
1
S
x
x
STD
21
5
0x08
TARGET_MODE
Target mode of block = Auto
unsigned8
1
S
x
x
STD
22
6
MODE_BLK
Set block mode
ds37
1
D
x
1
0x08
Actual
Actual mode
2
0x98
Permitted
Permitted modes: Auto, O/S, MAN
3
0x08
Normal
Set mode in normal operation
STD
23
7
ALARM_SUM
Alarm status
ds42
1
D
x
1
0x0000
Current_Alarm
Current Alarm
2
0x0000
Unacknowledged
Not acknowledged
3
0x0000
Unreported
Not reported
4
0x0000
Disabled
Disabled
STD
24
8
BATCH
Information for Batch-Process
ds67
1
S
x
x
1
BATCH_ID
2
RUP
3
OPERATION
4
PHASE
MAND
25
9
SP_D
ds34
1
D
x
x
0
Value
0x1B
Status
O(A),
M(B)
26
10
OUT_D
ds34
1
D
x
x
0
Value
0x4F
Status
OPT
28
12
READBACK_D
ds34
1
D
x
0
Value
0x4F
Status
O(A),
M(B)
30
14
RCAS_IN_D
ds34
1
D
x
x
0
Value
0x1B
Status
SPEC
0x0123
IN_CHANNEL
Pointer for input (FEEDBACK)
unsigned16
1
N
O(A),
M(B)
33
17
0x0122
CHANNEL
Pointer for output (POSITIONING_VALUE)
unsigned16
1
S
x
x
MAND
34
18
0
INVERT
unsigned8
1
S
x
x
O(A),
M(B)
35
19
30.0
FSAVE_TIME
float
1
S
x
x
O(A),
M(B)
36
20
1
FSAVE_TYPE
unsigned8
1
S
x
x
O(A),
M(B)
37
21
0
FSAVE_VAL_D
unsigned8
1
S
x
x
O(A),
M(B)
38
22
RCAS_OUT_D
ds34
1
D
x
0
Value