7. Appendix
7.1. Operation with SIMATIC PDM
Parameters can be set and measured values monitored using the SIMTIC PDM operating program and a PC
or programming unit.
You will find detailed information on the Internet at
http://www.feldgeraete.de/00/produkte/simatic_pdm/infos.html
7.2. Device Data Base File (GSD)
PROFIBUS DP / PA requires files for cyclic data transmission which you can download from the Internet at
German:
http://www.ad.siemens.de/csi/gsd
English:
http://www.ad.siemens.de/csi_e/gsd
(3 KB) PA independant language
(4 KB) DP independant language
(3 KB)
Bitmap_Device
7.3. Technical Data
PROFIBUS PA
(according to DIN IEC 1158-2)
Bus voltage
9 ... 32 V (not Exi)
9... 24 V option Exi
Current consumption
10 mA (typical)
Baud rate
31,25 kBit/s
Electrical isolation
PROFIBUS against other electronics or housing: 1810 VAC
PROFIBUS DP
(according to EN 50170)
Baud rates
9.6, 19.2, 93.75, 187.5, 500 kBit/s, 1.5, 3, 6,12 MBit/s
Electrical isolation
PROFIBUS against other electronics or housing: 50V=
PROFIBUS DP / PROFIBUS PA
C2-connections
4 connections to master class 2 are supported.
automatic connection clearing 60 s after communications breakdown.