ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
87
2.8.1
Installing the Optional ULTRAMAT / OXYMAT 6 / CALOMAT 6 Module and OXYMAT 61
☛
Note
2 PROFIBUS interfaces are required for the ULTRAMAT/OXYMAT or ULTRAMAT 6 E two channel
device.
Installation instructions:
•
Disconnect the device from the mains
•
Unscrew and remove the housing cover
•
Remove the old optional module if installed
•
Fit the module(s) into the rail on the base
•
Fasten the module(s) with the M3 screws to the rear panel between connectors
•
Connect the connectors to the motherboard
•
Fit the clamping bar
•
Screw on housing cover
•
Connect the device to the PROFIBUS
•
Connect the device to the mains
Connect 26-pin
ribbon cable
Rear panel
Optional modules
Motherboard
channel 1
Motherboard
channel 2
Fig. 2.1