System Overview
Troubleshooting
Troubleshooting Failed EQC
If Reader repeatedly fails Electronic QC, then this
indicates it is not fit for use. It is not possible to use
a Reader that has failed Electronic QC. Complete
the following troubleshooting steps if the message,
“Electronic QC failure,” displays on the Host.
1. Close the Reader screen
2. Turn the Reader OFF and back ON
3. If the Reader connects successfully to the Host and
passes EQC, it is acceptable for use. If EQC continues
to fail, contact Technical Support