![Siemens Healthcare epoc Resource Manual Download Page 24](http://html.mh-extra.com/html/siemens-healthcare/epoc/epoc_resource-manual_1266126024.webp)
24
System Overview
As Needed Maintenance
Replacing the Reader Battery
1. Remove the battery door screw using a Philips
#0 or 3/16” slotted screwdriver
2. Lift the battery door from the Reader
3. Remove the battery from its compartment
and unplug the battery from the Reader. Use
tweezers if necessary
4. Insert the new battery connector to the Reader.
The connector orientation tab should be visible
Battery door screw
Orientation tab
Note:
epoc Reader Battery Replacement SMN: 10736373