18
System Overview
Sample Processing
Reviewing Critical Results
1. After analysis is complete, any results outside of
critical range will appear in red with a critical-range
indicator. The Critical Actions button is displayed on
the right side of the Host screen
2. Select the
Critical Actions icon
to display
additional options (if configured)
3. Select the
Action drop down menu to select the
healthcare professional notified
4. Enter the name of the healthcare professional in
the
Notify field
5. Select
Yes or No if the results were read back to the
operator
6. If needed, select
Change to enter the appropriate
Time and Date
7. Select
Save
8. Remove the used test card and dispose in a
biohazard receptacle
9. Insert a new test card and repeat previous sample
processing steps if desired
10. When all testing with the Reader is complete, close
testing by selecting the Red X at the top right to
close the Reader Screen for that Reader.
Note:
After disconnecting all Readers from
running tests, test results can be sent to a
data manager by pressing the data manager
synchronization button.
Note:
The Administrator may configure the epoc
Host to synchronize upon closing a test. In
this configuration, synchronization with the
data manager occurs immediately after the
Reader Screen is closed at the end of a test.