System Overview
Sample Processing
Sample Injection (Syringe)
1. When ready for sample injection, the Host will display
the message
Inject sample on the Test Results page tab
2. Once sample is obtained, discard a drop of blood as
volume permits
3. Hold the syringe barrel vertically between fingertips
and thumb and perpendicular to the test card
4. Using slight downward pressure, secure the syringe’s
tip into the sample entry port of the test card. Rotate
the syringe clockwise ¼ turn to ensure a good seal
5. While maintaining downward pressure, use the index
finger of the other hand to steadily depress the syringe
plunger with one continuous motion until the Reader
provides an audible beep and the Test Status Indicator
flashes green indicating enough sample was received.
The Host will display the message
Analyzing sample
on the Test Results page tab