System Overview
Sample Processing
Reviewing Results
1. Once analysis is complete, test results will display on the
Test Results page tab
• If the prompt “To view results: Enter patient ID” appears, enter the
patient ID number. Select
Save
2. Select
Gases+ to view blood gas results
3. Select
Chem+ to view chemistry results
4. Select
Meta+ to view metabolite results
5. If a test was not selected prior to sample analysis and is needed, select
Test Selection; select the check box for the desired test, select Save
6. If needed, select the test information page tab to enter additional
patient information, select
Save
7. Select the Additional Documentation icon
(if configured) to display
additional options
8. Remove the used test card and dispose in a biohazard receptacle
9. Insert a new test card and repeat previous sample processing steps
if desired
10. When all testing with the Reader is complete, close testing by selecting
the Red X at the top right to close the Reader Screen for that Reader.
Note:
Sample analysis takes
approximately 35
seconds to complete
Note:
After disconnecting all Readers from running
tests, test results can be sent to a data manager by
pressing the data manager synchronization button.
Note:
The Administrator may configure the epoc Host to synchronize upon closing
a test. In this configuration, synchronization with the data manager occurs
immediately after the Reader Screen is closed at the end of a test.