3. Remove the contents of the package.
Inspecting the Contents of the Foil Package
The foil package includes the contents displayed below
:
Figure 6 DCA Systems Hemoglobin A
1c
Reagent Cartridge Package
Discard the reagent cartridge if any of the following conditions exist:
•
The cartridge is damaged.
•
The flexible pull-tab is loose or missing.
•
The desiccant bag is missing or open.
•
Loose desiccant particles are found inside the foil package.
•
If the foil package is open for more than 60 minutes.
CAUTION
When handling the reagent cartridge, do not touch or otherwise
contaminate the optical window or erroneous test results may
occur.
1
Flexible Pull Tab
2
Reagent Cartridge
3
Desiccant bag
4
Optical Window