DCA Vantage Operator’s Guide
81
Calibration
Calibrating for Microalbumin/Creatinine
The calibration card for the Microalbumin/Creatinine reagent cartridges has
data on two sides. You must scan both sides of the calibration card for the lot of
reagent cartridges before you can use them on the DCA Vantage system.
1. Locate the dot on next to the barcode track.
2. Locate the barcode on one side of the Calibration card.
3. Hold the card so that the barcode faces to the right.
4. Insert the calibration card into the top of the barcode track.
Figure 26 Scanning Microalbumin/Creatinine calibration card
5. Hold the calibration card gently against the right side of the track and
smoothly slide the card down.
A beep sounds to signal a successful scan.
If no beep sounds, repeat the scanning procedure. If you repeatedly fail to
hear a beep, refer to
Troubleshooting
‚ page 107.
6. Locate the barcode on the second side of the calibration card.
7. Repeat steps 3 – 5.
8. To return to the Home screen, select
OK
.