Comparing New Values with Initial Values
Compare the new values obtained with the results recorded initially for your
system. The Mean Transmittance should be within the range of 0.9500 to
1.0500, and should not vary more than plus or minus 0.0100. The Standard
Deviation should be less than 0.0015 and the Drift should be less than
0.0140.
1. Locate the barcode on the optical test cartridge.
2. Hold the cartridge so that the barcode faces right.
3. Insert the cartridge into the barcode track.
4. Quickly and smoothly slide the cartridge down the barcode track.
A beep sounds to signal a successful scan.
Figure 36 Scanning the optical test cartridge