Result details
The HbA
1c
Test displays the Hemoglobin A
1c
percent. The reporting range for
patient samples is 2.5% to 14.0% HbA
1c
and for control samples is 2.5% to
16.0% HbA
1c
. The test is linear throughout the range.
A result preceded by a less than (<) sign in the display indicates that the
concentration is below the lower limit of the test, that is <2.5%.
A result preceded by a greater than (>) sign in the display indicates that the
concentration is above the upper limit of the test, that is >14.0% for patient
samples and >16.0% for control samples.
If you set the Reference Range for the HbA
1c
Patient Test at 3.5% to 12%, a
value of 3.2% is displayed as "3.2% -" showing that it is within the Reporting
Range but below the Reference Range. Similarly a value of 12.5% is displayed
as "12.5% +" showing that it is within the Reporting Range but above the
Reference Range. A value of 5.2% is displayed as "5.2%" because it is within
both the Reference and Reporting Ranges.
NOTE:
You can print, enter, or edit the sample data while the results are
displayed.
To print the results, select
Next
.
NOTE:
The results only print automatically when the system is set to autoprint.
For more information, see
Printers Menu
‚ page 68.
To print manually, select
.
NOTE:
If the print button is not enabled, then the printer is disabled in setup, or
there is no paper.
Summary of Contents for DCA Vantage
Page 1: ...From Wilburn Medical USA 1 877 WILBURN 945 2876 www wilburnmedicalusa com ...
Page 3: ...DCA Vantage Operator s Guide ...
Page 15: ...14 DCA Vantage Operator s Guide ...
Page 123: ...122 DCA Vantage Operator s Guide ...
Page 125: ...124 DCA Vantage Operator s Guide File Management ...
Page 131: ...130 DCA Vantage Operator s Guide System Configuration ...
Page 161: ...160 DCA Vantage Operator s Guide ...