DCA Vantage Operator’s Guide
Tro
ublesh
oot
in
g
E31 - Cartridge
temperature is
very low
The temperature
measured by one of the
cartridge holder
thermistors is <(T
setpoint
-
12.9)
º
C while the
system is idle, during an
Optical Test, during the
first 5 minutes of an
HbA
1c
test, or the first 5
minutes of a
Microalbumin/Creatinine
test.
1. Discard the sample.
2. Contact your technical support
provider.
E32 - Cartridge
temperature
too low during
measurement
The temperature
measured by one of the
cartridge holder
thermistors is
≤
(T
setpoint
-
0.9)
º
C during critical
measurements of a
sample test sequence.
1. Discard the sample.
2. Contact your technical support
provider.
E33 - Cartridge
temperature
too high during
measurement
The temperature
measured by one of the
cartridge holder
thermistors is
≥
(T
setpoint
+
1.1)
º
C during critical
measurements of a
sample test sequence.
1. Discard the sample.
2. Contact your technical support
provider.
E34 - Cartridge
temperature
very high
The temperature
measure by one of the
cartridge holder
thermistors is
≥
(T
setpoint
+
2.1)
º
C.
1. Discard the sample.
2. Contact your technical support
provider.
E35 - Thermal
control system
error - high
The temperature
measure by one of the
cartridge holder
thermistors is
≥
50
º
C.
1. Discard the sample.
2. Restart the system.
3. If you are unable to perform a
successful restart, contact
your local technical support
provider.
Error Code
Description
Remedy
Summary of Contents for DCA Vantage
Page 1: ...From Wilburn Medical USA 1 877 WILBURN 945 2876 www wilburnmedicalusa com ...
Page 3: ...DCA Vantage Operator s Guide ...
Page 15: ...14 DCA Vantage Operator s Guide ...
Page 123: ...122 DCA Vantage Operator s Guide ...
Page 125: ...124 DCA Vantage Operator s Guide File Management ...
Page 131: ...130 DCA Vantage Operator s Guide System Configuration ...
Page 161: ...160 DCA Vantage Operator s Guide ...