MAINTENANCE
10
10.4
Maintenance intervals
System compo-
nents
Description
Intervals
1x/
week
1x/
month
1x/
year
Gas sampling
probe (option)
Check filters and seals.
Replace filters regularly depending on wear.
Note:
See the Operating Instructions of the gas sampling probe.
X
Sample gas line
(option)
Check heating function. Clean.
Note:
Observe the operating instructions/assembly guidelines for fit-
ting heating hoses.
X
Filter (option)
Check condition. Degree of contamination different depending on
the application. Check for condensate in filter housing bottom. Drain
off condensate when necessary.
Note:
Observe Operating Instructions/Data Sheets.
X
Glass wool filter
(option)
Check condition. Wear different depending on the application.
Note:
When replacing brass or glass wool, make sure no particles
can penetrate the control cabinet.
For 2/3 dark coloring
and/or heavy degradation,
replace filter
Gas washer
(option)
Function check of the gas washer with automatic media replace-
ment. Inlet/outlet must be free.
X
Humidifier con-
tainer (option)
Check function.
Check liquid reserves (solution and/or distilled water) of humidifier
container and top up if necessary. Filling quantity approx. 50% of
container volume. Check for contamination / blockage of the con-
tainers, bubbling frits and Raschig rings. Clean as required.
Note:
Bubbling frits / Raschig rings are fragile.
X
Sample gas pump
(option)
Check function.
Note:
Observe relevant Operating Instructions.
X
Water trap
Replace.
Note:
Condensate in the water trap blocks it. Thus, the gas flow is
interrupted and a “System error” is signaled.
X
Cabinet vent
Filter fan
(option)
Function check
Clean heavily contaminated filter pads and replace if necessary.
Note:
Replace/clean filters in good time.
Clean filter pads by flushing or blowing.
Note:
Observe the filter fan Operating Instructions.
X
Cabinet
thermostat
(option)
Check function in connection with the filter fan.
X
Condensate
collector
(with level switch)
Check container state. Empty when necessary. Observe the safety
instructions.
Note:
Sample gas condensate is hazardous waste. Observe
regulations on condensate disposal according to the Water
Resources Law (WHG).
Caution
: Serious damage to environment, nature and health by toxic
condensate.
X