ADJUSTING
8
8.6.1
Clearing malfunctions
8.6.2
Resetting the maintenance request after automatic adjustment abort
▸
In case of an adjustment abort, the adjustment values are discarded.
▸
A maintenance request is activated.
▸
Turn key switch to analyzer adjustment and back again (acknowledgment).
▸
The maintenance request is reset.
8.6.3
Manual adjustment abort on the analyzer
▸
After a manual adjustment abort, the adjustment values are discarded.
▸
Maintenance request is not activated.
Malfunction
Possible cause
Action
Adjustment
abort
Test gas container empty
Replace test gas container
Test gas container not connected
Connect test gas container
Test gas path clogged
Clean test gas path
Test gas path interrupted
Connect test gas path
Test gas pressure too low
Set test gas pressure to 1.3 bar
Non-return valve on the probe does
not open
Check non-return valve, replace if
necessary