25
8019897/V2-0/2017-12| SICK
O P E R A T I N G I N S T R U C T I O N S | PowerCEMS100
Subject to change without notice
INSTALLATION
5
5
Installation
5.1
Preparing the installation location
▸
Observe ambient conditions,
(see “Technical data”, page 64)
.
– Under a roof
– Protection against direct sun/heat radiation
– Heavy dust load
– Corrosive atmosphere
▸
System cabinet dimensions (see Technical Data / Technical System Documentation).
▸
Ensure floor carrying capacity.
▸
Set up the system cabinet in a low-vibration environment.
Vibrations can influence measurement, plan vibration damping.
▸
Set up the system cabinet as close as possible to the sampling point.
– Short sample gas lines mean short T-90 times.
– Maximum length of sample gas line: 70 m.
▸
Provide a suitable installation location for test gas cylinders (only when used).
▸
Air outlet of the system control cabinet ventilation / climate control
– Do not block air outlet of the system control cabinet ventilation and climate control.
▸
Provide (individual) attachments for the system cabinet.
▸
For installation on gratings, provide a base plate under the system cabinet. As protection
against falling parts or liquids (condensate).
Note
Connection to the gas supply may only be performed by skilled persons or
qualified personnel who, based on their technical training and knowledge as
well as knowledge of the relevant regulations, can assess and perform the tasks
given and recognize and avoid hazards.
Note
The operator is responsible for preparing the installation location.
Note
Observe the assembly guidelines for fitting heating hoses.
Note
▸
Observe local regulations on installing gas cylinders.
▸
Provide a suitable installation location for the pressure control unit.
WARNING
ACIDIC OR ALKALINE CONDENSATE
Serious injuries to the skin and serious environmental damage.
▸
Wear gloves and protective clothes.
▸
Collect dripping condensate.
▸
Dispose of condensate according to valid local environment protec-
tion guidelines.