CLEARING MALFUNCTIONS
Gas com-
ponent
Bad Config. (text)
F
ailure
Error in calculation models.
Please contact SICK Customer Service.
File I/O (text)
Error in file system.
Restart system.
If the error continues to exist:
Please contact SICK Customer Service.
Measurement
range x
X
tended Current measuring range x ( x = 1 .. 8)
---
Measurement
value out of range
U
ncertain Measured value outside calibration range.
Check measured values for plausibility.
Measurement
value range
warning
X
tended Measurement beyond a warning threshold
defined during calibration.
Medium pressure
out of range
U
ncertain Sample gas pressure outside calibrated range.
Check sample gas pressure.
Medium pressure
warning
X
tended Sample gas pressure beyond warning threshold.
Medium tempera-
ture out of range
U
ncertain Sample gas temperature outside calibrated
range.
Check sample gas temperature.
Medium
temperature
warning
X
tended Sample gas temperature beyond warning
threshold.
Absorption range
warning
X
tended Absorption in active measuring path above
warning threshold.
Standard setting of warning threshold: 1.8
extinction units.
Check:
- Window dirty
see “Cleaning the window”, page
- Dust content in sample gas too high?
- Sample gas concentration too high?
Absorption out of
range
F
ailure
Absorption in active measuring path too high.
Standard setting of error threshold: 2 extinction
units.
Syntax error
Error in concentration calculation.
Please contact SICK Customer Service.
Processing error
Numerical
(DivZero)
Numerical error in concentration calculation.
Numerical
(IppError)
Numerical
(MatSing)
OS error (text)
Error in operating system.
Restart system.
If the error continues to exist:
Please contact SICK Customer Service.
Spectr. resolution
out of range
Resolution of spectrometer wrong.
Please contact SICK Customer Service.
Spectral
evaluation
U
ncertain Error in calculation of spectra.
[1] System = SR-unit
Probe = probe
P = pressure sensor
T = temperature sensor
Gas component
[2] This Table also contains recommended solutions that can only be performed by specially trained personnel.
Initia-
tor
[1]
Text
Classifi-
cation
Description
Possible causes/remedy
[2]
Table 18
Error messages