CLEARING MALFUNCTIONS
7
7.4
Inadequate purge air supply (for GMP probe)
Indications of inadequate purge air supply
●
Unusual noise from the area of the purge air unit.
●
On systems with pressure difference sensor: An appropriate error message occurs.
●
Rise in enclosure temperature.
●
Unusually rapid contamination of the GM32 window.
Check the purge air unit
▸
Remove the purge air hose on the SR-unit: A strong air flow must be noticeable.
▸
Reinstall the purge air hose immediately.
Measures if purge air supply is inadequate
▸
If the purge air unit is not immediately functional: Remove the SR-unit from the gas duct
(swiveling out is sufficient for short malfunctions).
▸
Immediately restore correct operation of the purge air unit or replace provisionally by a
different purge air supply with at least the same purge air throughput.
Information for fast correction of faults
●
Air filter of purge air unit clogged?
●
Purge air hose slipped off or broken?
●
Power supply of purge air unit failed?
7.5
Malfunctions on the connection unit
A green LED lights on each power supply unit in the connection unit.
If no LED lights: Check the voltage supply of the connection unit.
Otherwise, please contact SICK Customer Service.
NOTE: Inadequate purge air supply can damage the gas analyzer.
▸
If there are signs of incorrect purge air supply, immediately perform the measures
listed below.