45
8012707/ZVF9/V2-1/2019-04 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32
Subject to change without notice
OPERATION
5
5.2.5.2
Check cycle
Results of the last check cycle.
Fig. 27: Check cycle
5.2.5.3
Alignment check (automatic optical alignment check; option)
This menu serves to view the values of automatic optical alignment.
▸
Arrow buttons: Toggle between “deviation” and “performed steps of tracking mirror”
displays.
▸
To exit the Menu item: Press the “
Back
” button.
5.2.5.4
A
djustments
Alignment adjustment (manual optical alignment)
Duct displayed
(In the example: Duct 1 of 3;
according to configuration)
Gas component
Zero adjust result
Absolute value (in unit configured)
Deviation ín percentage of FS
1
Span adjust result (70%)
Absolute value (in unit configured)
Deviation ín percentage of FS
1
Date and time
of last
Check cycle
1
FS = full scale value:
Scale limit value of the assigned analog output
▸
Perform this check only when the SR-unit is at operating temperature (in operation
for at least 30 minutes).
▸
Automatic mirror adjustment - do not perform manual adjustments.
Further information
▸
Perform this work only when the SR-unit is at operating temperature (in operation for
at least 30 minutes).